Glucose level | Criteria | Number | % within |
---|---|---|---|
Criteria based on Chinese consensus guidelines for hospital blood glucose monitoring | |||
<5.5 mmol/L | 95% within ±0.83 mmol/L | 99 | 97.0 (96/99) |
≥5.5 mmol/L | 95% within ±15% | 48 | 95.8 (46/48) |
Criteria based on CLSI POCT 12-A3 system accuracy for hospital blood glucose monitoring | |||
≤5.55 mmol/L | 95% within ±0.67 mmol/L | 102 | 95.1 (97/102) |
>5.55 mmol/L | 95% within ±12.5% | 45 | 95.6 (43/45/) |
4.2 mmol/L | 98% within ±0.83 mmol/L | 147 | 98.0 (144/147) |
≤4.2 mmol/L and within ±20% | |||
>4.2 mmol/L |